MAK 3003 EASI Ultrasampler Controller Type 6970-31, Configuration, Software version EASI 1.00, SAK 200311
4.3.6
Motor 1
Motor 1
Operating mode
Not installed
Single sample
Tour sample
Vial Volume
Content of sample bottle in ml
Overfill Tolerance
Quantity permitted in the sample bottles expressed as a
percentage
e.g. 120%
Bottle may be filled up to 20% over the rated volume
Start Delay
Delay from the milk sensor signal ("milk present") up to sampling start
[s].
If this value is reached, the fat content gets higher for all quantities.
Calibration Factor
The pumping hose calibration factor indicates how many µl are filled
per sampler rotation. (Enter 270 µl. Change the value if sampling vial
filling volume deviates from the set value.)
Pre Flushing From Estimate
In case of intake quantities (set quantities) lower than the value en-
tered here, no pre-flushing occurs. [L]
Flushing Steps
Number of flushing steps when milk is detected in the pipe.
Minimum Pulse Steps
Minimum number of motor steps.
(10)
Minimum Pulse Frequency
Minimum rpm of the sampler. If this rpm value is not reached, this
triggers intermittent mode.
(50)
4.3.7
Motor 2
Motor 2
Operating mode
Not installed
Single sample
Tour sample
Vial Volume
Content of sample bottle in ml
Overfill Tolerance
Quantity permitted in the sample bottles expressed as a
percentage
e.g. 120%
Bottle may be filled up to 20% over the rated volume
Start Delay
Delay from the milk sensor signal ("milk present") up to sampling start
[s].
If this value is reached, the fat content gets higher for all quantities.
Содержание MAK 3003