Copyright 2013 Baker Hughes Company.
34 | Dry
TC / Liquid
TC Temperature Calibrators–English
2. Press the
or
key to select the
parameter.
Figure
43: Parameter Level SLoF (Cooling Gradient)
appears on the upper display.
flashes on the lower display.
3. Press the
key.
Figure
44: Cooling Gradient (SLoF) Entry
flashes on the upper display.
The current cooling gradient appears on the lower display.
4. Press the
key to increase the cooling gradient.
5. Press the
key to decrease the cooling gradient.
Note:
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys
are held pressed for at least one second, the value increases or decreases quickly and after
two seconds even more quickly; this means the desired value can be reached rapidly.
6. Press the
key to confirm the cooling gradient.
The display returns to the parameter level.
Note:
If no key is pressed for approximately 15
seconds, the controller automatically returns to
the calibration mode.
Note:
See Section
5.4.4.5 for details on new cooling gradient parameter activation.
5.4.4.4 Setting the Duration Time
Duration time
is active once set point SP1 has been reached. After
has
expired, the controller will regulated to set point SP2.
The duration time is adjustable from 99:99
[hh:mm] to 00:00
[hh:mm].
Note:
If
=
(in no function), duration time has been disabled.
Содержание Druck Dry TC 165
Страница 1: ...Druck com Dry TC 165 Dry TC 650 Liquid TC 165 Liquid TC 255 Temperature Calibrators Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2013 Baker Hughes Company iv Dry TC Liquid TC Temperature Calibrators English...
Страница 56: ...Copyright 2013 Baker Hughes Company 48 Dry TC Liquid TC Temperature Calibrators English...
Страница 57: ......