Copyright 2013 Baker Hughes Company.
28 | Dry
TC / Liquid
TC Temperature Calibrators–English
5.4.3 Set Point Memory
Up to four set point temperatures can be stored and retrieved from the set point memory.
5.4.3.1 Saving Set Point Temperatures
1. Press the
key when in calibration mode for approximately 5
seconds. The main menu
opens.
Figure
30: Operator (OPEr) Menu
(operator) appears on the upper display.
LED SET flashes on the lower display.
2. Press the
key again. The group level opens.
Figure
31: Group Level SP (Set Point)
appears on the upper display.
(set point) appears on the lower display and LED SET flashes.
3. Press the
key again. The parameter level opens.
Figure
32: Parameter Level SP1 (Set Point 1)
appears on the upper display.
(set point 1) appears on the lower display and LED SET flashes.
4. Use the
or
key to select one of the four set point memories SP1, SP2, SP3 or SP4.
Содержание Druck Dry TC 165