Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 35
1. Enter the gradient control parameter level. See Section
Figure
45: Parameter Level SLor (Heating Gradient)
appears on the upper display.
flashes on the lower display.
2. Press the
or
key to select the
parameter.
Figure
46: Parameter Level dur.t (Duration Time)
appears on the upper display.
flashes on the lower display.
3. Press the
key.
Figure
47: Duration Time (dur.t) Entry
flashes on the upper display.
The current duration time appears on the lower display.
4. Press the
key to increase the duration time.
5. Press the
key to decrease the duration time.
Note:
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys
are held pressed for at least one second, the value increases or decreases quickly and after
two seconds even more quickly; this means the desired value can be reached rapidly.
6. Press the
key to confirm the duration time.
The display returns to the parameter level.
Note:
If no key is pressed for approximately 15
seconds, the controller automatically returns to
the calibration mode.