Copyright 2013 Baker Hughes Company.
16 | Dry
TC / Liquid
TC Temperature Calibrators–English
The maximum fill levels are shown below:
Figure
12: Maximum Fill Level
Note:
The maximum fill level line for the micro bath tank is located next to the upper edge of the
aluminum lining. The maximum fill level line for the tub insert is below the sleeve exchange tool
fixture.
4.3.5.9 Filling with Calibration Liquid
Figure
13: Filling with Calibration Liquid
Micro bath filling procedure:
1. Unscrew the transport cover of the micro bath.
1 Micro bath tank.
2 Tub insert.
3 Sensor cage.
4 Tub insert transport cover.
5 Magnetic stirrer.
Micro Bath Tank
Tub Insert
Maximum Fill Level (h)
150
mm (5.91″)
136
mm (5.35″)
Maximum Fill Volume
~0.45
litres (15.22
fl
oz)
~0.32
litres (10.82
fl
oz)
INFORMATION
Pay attention to maximum calibration liquid filling level, see
When filling, leave enough room for thermal expansion, displacement by the
DUT and level rise due to stirring.
h
4
3
5
1
2
h