Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 9
4.3.1.1 Installation
The adaptor sleeve is inserted into the dry block using the insert exchange tool, see Figure
item
3.
4.3.1.2 Removal and Cleaning
1. Allow the calibrator cool before the adaptor sleeve is removed.
2. Pull the adaptor sleeve out of the dry block with the help of the insert exchange tool, see
3 item
3.
3. Clean the adaptor sleeve and the dry block to prevent inserts from getting stuck in the dry
block.
4.3.2 Air Shield Insert
The air shield insert provides optimum radial and axial temperature distribution. The borehole
divider provides a flexible and cost-effective adaptation to various calibration tasks. For air shield
insert compatibility, see Section
6.
Figure
4: Air Shield Insert
4.3.2.1 Installation
The air shield is inserted into the dry block using the insert exchange tool, see Figure
4 item
2.
4.3.2.2 Removal and Cleaning
1. Allow the calibrator cool before the air shield is removed.
2. Pull the air shield out of the dry block with the help of the insert exchange tool, see Figure
item
2.
3. Clean the air shield and the dry block to prevent inserts from getting stuck in the dry block.
INFORMATION
Use only Druck adaptor sleeves.
If in doubt, contact Druck.
1 Air shield insert.
2 Insert exchange tool.
2
1
Содержание Druck Dry TC 165
Страница 1: ...Druck com Dry TC 165 Dry TC 650 Liquid TC 165 Liquid TC 255 Temperature Calibrators Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2013 Baker Hughes Company iv Dry TC Liquid TC Temperature Calibrators English...
Страница 56: ...Copyright 2013 Baker Hughes Company 48 Dry TC Liquid TC Temperature Calibrators English...
Страница 57: ......