Copyright 2011 Baker Hughes Company.
44 | DPS 8000 Instruction Manual–English
Appendix
A.
CANopen Object Dictionary
A.3.7 Object 6123h: AI Input Scaling 2 PV
This object defines the process value of the second calibration point for the analogue input
channel. It is scaled in physical unit of process value.
OBJECT DESCRIPTION
INDEX
6123
Name
AI Input scaling 2 PV
Object Code
Array
Data Type
REAL32
ENTRY DESCRIPTION
Sub-Index
000
Description
number of entries
Access
RO
PDO Mapping
No
Default Value
0.0
Sub-Index
001
Description
AI Input scaling 2 PV 1
Data Type
REAL32
Access
RW
PDO Mapping
No
Default Value
1.0
Sub-Index
002
Description
AI Input scaling 2 PV 2
Data Type
REAL32
Access
RW
PDO Mapping
No
Default Value
1.0
A.3.8 Object 6124h: AI Input Offset
This object defines the additional offset value for the analogue input channel. It is scaled in
physical unit of process value.
OBJECT DESCRIPTION
INDEX
6124
Name
AI Input offset
Object Code
Array
Data Type
REAL32
ENTRY DESCRIPTION
Sub-Index
000
Description
number of entries
Access
RO
PDO Mapping
No
Default Value
0.0
Содержание Druck DPS 8000 Series
Страница 1: ...Druck com DPS 8000 Series CAN bus Pressure Transducer Instruction Manual ...
Страница 2: ......
Страница 6: ...Copyright 2011 Baker Hughes Company iv DPS 8000 Instruction Manual English ...
Страница 10: ...Copyright 2011 Baker Hughes Company viii DPS 8000 Instruction Manual English ...
Страница 28: ...Copyright 2011 Baker Hughes Company 18 DPS 8000 Instruction Manual English ...
Страница 62: ...Copyright 2011 Baker Hughes Company 52 DPS 8000 Instruction Manual English Appendix C Bibliography ...
Страница 63: ......