Copyright 2011 Baker Hughes Company.
English–DPS 8000 Instruction Manual | 9
3.7 Returned Goods Procedure
To repair or calibrate the sensor, return it to the applicable Druck Service Department.
Please contact our Service Department, and get a Return Authorization number.
Please supply these details:
•
Product (e.g. TERPS
820C Pressure Sensor)
•
Pressure range
•
Serial number
•
Details of defect / work to be undertaken
•
Calibration traceability requirements
•
Operating conditions
3.7.1 Safety Precautions
To prevent possible injury when we receive the product, you must also tell us if the product has
been in contact with hazardous or toxic materials. Please supply the applicable Control of
Substances Hazardous to Health (COSHH) or Material Safety Data Sheet (MSDS) references
and precautions.
3.7.2 Important Notice
Service or calibration by unauthorized sources will affect the warranty and may not guarantee
further performance.
3.8 Electromagnetic Compatibility
The pressure sensor complies with the European Electromagnetic Compatibility Directive
2014/30/EU.
When appropriately installed the sensors meet and exceed the Commercial and Industrial
specifications indicated in Table
3.8.1 Power Supply and Metering
The quality of the power supply and monitoring equipment will directly affect the EMC
performance of the entire system. Since “Druck Limited” has no control over the installation of the
sensor it must remain the responsibility of the user to ensure that the EMC performance of the
system is adequate.
To maintain good immunity from electromagnetic disturbances present on the system power
supply, the power supply should filter any transient interference from the incoming line and
present a clean regulated DC supply to the sensor. The monitoring equipment should likewise be
immune from the effects of electromagnetic disturbances and not impart disruptive signals on the
connections to the sensor.
The sensor is not intended for connection to a DC distribution network.
Table
5: EMC Standards
TERPS 8*00 Series
Models DPS 8***-T*-A*-C*-**-**[-…]
EN 61000-6-1:2007
EN 61000-6-2:2005
EN 61000-6-3:2007 + A1:2011
EN 61000-6-4:2007 + A1:2011
EN 61326-1:2013
EN 61326-2-3:2013
Содержание Druck DPS 8000 Series
Страница 1: ...Druck com DPS 8000 Series CAN bus Pressure Transducer Instruction Manual ...
Страница 2: ......
Страница 6: ...Copyright 2011 Baker Hughes Company iv DPS 8000 Instruction Manual English ...
Страница 10: ...Copyright 2011 Baker Hughes Company viii DPS 8000 Instruction Manual English ...
Страница 28: ...Copyright 2011 Baker Hughes Company 18 DPS 8000 Instruction Manual English ...
Страница 62: ...Copyright 2011 Baker Hughes Company 52 DPS 8000 Instruction Manual English Appendix C Bibliography ...
Страница 63: ......