Copyright 2011 Baker Hughes Company.
22 | DPS 8000 Instruction Manual–English
Appendix
A.
CANopen Object Dictionary
ENTRY DESCRIPTION
Access
RW
PDO Mapping
No
Default Value
0x0
A.1.9 Object 1010h: Store Parameter Field
This entry supports saving of parameters in non volatile memory. With a read access the device
provides information about its saving capabilities.
Several parameter groups are distinguished:
For saving the signature “save” (0x65766173) must be written.
OBJECT DESCRIPTION
INDEX
1010
Name
Store Parameter Field
Object Code
Array
Data Type
UNSIGNED32
ENTRY DESCRIPTION
Sub-Index
000
Description
Number of Entries
Access
RO
PDO Mapping
No
Default Value
0x1
Sub-Index
001
Description
Save all Parameters
Data Type
UNSIGNED32
Access
RW
PDO Mapping
No
Default Value
0x1
A.1.10 Object 1011h: Restore Default Parameters
This entry supports restoring of default parameters. With a read access the device provides
information about its capabilities to restore these values.
Sub-Index
Parameter Groups
1
All parameters.
2
Communication parameters.
3
Application parameters.
4 to 127
Manufacturer defined parameters.
Содержание Druck DPS 8000 Series
Страница 1: ...Druck com DPS 8000 Series CAN bus Pressure Transducer Instruction Manual ...
Страница 2: ......
Страница 6: ...Copyright 2011 Baker Hughes Company iv DPS 8000 Instruction Manual English ...
Страница 10: ...Copyright 2011 Baker Hughes Company viii DPS 8000 Instruction Manual English ...
Страница 28: ...Copyright 2011 Baker Hughes Company 18 DPS 8000 Instruction Manual English ...
Страница 62: ...Copyright 2011 Baker Hughes Company 52 DPS 8000 Instruction Manual English Appendix C Bibliography ...
Страница 63: ......