REPAIR AND REPLACEMENT PROCEDURES
OPTIONAL EQUIPMENT REMOVAL AND INSTALLATION PROCEDURES
WBPEEUI110505A0
8 - 11
EZ CAL Option Removal
1. Use the flat-blade screwdriver to remove the mounting
screw that secures the EZ CAL option to the electronics
housing.
2. Remove the EZ CAL option.
EZ CAL Option Installation
1. Position the EZ CAL option on the electronics housing as
shown in Figure
.
2. Use the flat-blade screwdriver to install and tighten the
mounting screw.
Flange Adapter Removal
1. Remove the transmitter from the process.
2. Using the adjustable end wrench, remove the flange
adapter bolts.
3. Remove the flange adapter O-ring.
Flange Adapter Installation
1. Refer to
and determine the proper flange adapter
O-ring for the application.
2. Coat the flange adapter O-ring with a film of Dow Corning
No. 4 compound and install.
3. Install the flange adapter.
4. Torque the flange adapter bolts to 45 to 49 Nm (33 to
37 ft-lbs).
NOTE: If the O-rings are PTFE, tighten to the same torque again
after
five days. This is necessary due to the cold flow properties of
PTFE. Tests have shown that a loss of torque of as much as
40 percent occurs over the five days. After a second tightening, no
loss of torque is apparent.
5. Install the transmitter on the process.
Vent/Drain Plug and Pipe Plug Removal
1. Remove the transmitter from the process.
2. Using the adjustable end wrench, remove the vent/drain
plugs and pipe plugs.