![Bailey Platinum standard Series Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/bailey/platinum-standard-series/platinum-standard-series_product-instruction_467169078.webp)
REPAIR AND REPLACEMENT PROCEDURES
COMPONENT REMOVAL AND INSTALLATION PROCEDURES
WBPEEUI110505A0
8 - 3
6. Use the 4-mm Allen wrench to remove the cell locking
screw (Fig.
7. Remove the cell lock (Fig.
8. Carefully remove the cell and process flange assembly,
feeding the cell cable and characterization board through the
opening left in the electronics housing by the cell.
9. Use the adjustable end wrench to remove all four of the
process flange nuts on the low pressure (-) side of the cell
(Fig.
).
10. Separate the flange from the cell, taking care not to dam-
age the isolating diaphragm (Fig.
).
Figure 8-1. Internal Transmitter Components
T 02 2 25 A
IN D IC ATO R
L O C K O U T
A /B
J 1
IN D IC ATO R
S C R E W S (4)
C H A R AC T E R IZ ATIO N
B O A R D S C R E W S (2)
C H A R AC T E R IZ ATIO N
B O A R D C O N N E C TO R
A M P LIFIE R
A S S E M B LY
S C R E W S (3)
B a iley
B a iley
Содержание Platinum standard Series