9.
TROUBLESHOOTING
9.1.
KNOWN ISSUES
Hardware
•
Web camera connected to the PC via USB 2.0 or USB 3.2 port could malfunction the sys-
tem.
•
Do not use USB 2.0 or USB 3.2 portable dongles for mobile Internet.
9.2.
ERRORS AND CAUSES
Error description
Possible causes
Remedy
1
The scanner doesn't turn
on by using the front
On\Off switch
•
Main connecting cable discon-
nected
•
Connect the main cable
•
Rear AC On\Off switch off
•
Turn On the rear AC switch
•
Fuses blow
•
Replace the fuses
•
If none of the above solves the
problem, contact E.G.S. srl cus-
tomer service
2
The scanner is On but
nothing is visible on the
monitor
•
DVI-I cable non properly con-
nected to the DVI-I port in the
scanner
•
Use the right DVI-I port in the
scanner
•
DVI-I cable not working
•
Change the cable or use the
ones provided in accessory box
•
A DVI-I to VGA adapter is present
in the DVI-I port
•
Remove the adapter and use
the cable provided in accessory
box
•
If none of the above solves the
problem, unplug the scanner
and contact the E.G.S. srl cus-
tomer service
3
The scanner is On but
warm up sequence gets
stucks
or
The scanner is On but no
warm up sequence
•
Projector failure
•
Contact E.G.S. srl customer ser-
vice
4
The
B9 Scan
application
doesn't run, and an error
•
The PenDrive is not connected
•
Connect the PenDrive in a scan-
ner USB 3.0 port
Содержание B9 Scan 500
Страница 1: ...B9 Scan 500 Operating Manual...
Страница 31: ...B9 Scan 500 Operating Manual Page31...
Страница 45: ...B9 Scan 500 Operating Manual Page45 Specifications B9 Scan 500 Package weight 24 Kg...
Страница 47: ...B9 Scan 500 Operating Manual Page47 Model Company Product Label B9 Scan 500 14 3 DIMENSIONS...
Страница 48: ...B9 Scan 500 Operating Manual Page48 14 4 ELECTRICAL CIRCUIT DIAGRAM...
Страница 49: ...B9 Scan 500 Operating Manual Page49 15 OPERATIONS CHECK LIST Putting into operations...
Страница 54: ...B9 Scan 500 Operating Manual Page54 KEYWORD INDEX No index entries found...