Safety
© b2 electronic GmbH
DHV0086 Rev06
9
DANGER
Authorized Personnel Only
The test area must be secured to keep non-qualified personnel off the
premises!
• Signs must warn all persons of the high voltage test area.
•
Only qualified electrical technicians should have access to the test
area.
•
Other persons must be accompanied by qualified electrical technicians
and must be informed of the risks involved.
WARNING
Radiation Hazard
Testing vacuum bottles, above their rated voltage, with DC can produce
dangerous X-rays.
NOTICE
Equipment Handling
DUT must have clean connections.
Testing instruments must only be repaired or modified by authorized
b2 personnel.
NOTICE
If required according to local safety regulations
Wear high voltage gloves when handling high voltage cables and
equipment.
WARNING
This is a class A product. In a domestic environment, this product may
cause technical interference, in which case the user may be required to
take adequate measures.