Design and Construction
© b2 electronic GmbH
DHV0086 Rev06
32
4�4 Operation Modes
Described below are the various HVA operation modes.
Test Modes, Output Modes (Waveform), Arc Management Modes, and Data
Transfer Modes.
Test Modes
The HVA can be operated in manual or automatic mode. For detailed procedure,
see 5.2 Manual Test Mode on page 40
, and
see 5.3 Automatic Test Mode
on page 55
.
Test mode
Characteristics
Manual
Designed to facilitate rapid testing. Test parameters of the last manual
test appear as the default settings.
• Test parameters can be changed before activating a test.
• Test types: VLF, VLF TD, DC, ST, SFL, VB
Automatic
Designed for testing with a predefined configuration in order to satisfy
specific requirements (e.g. IEEE or IEC standards).
•
The test sequence must be configured and saved before testing.
• Test types: VLF, VLF TD, DC, ST, VB