62
User Manual v2.0
MAG
MENU
COUNTERS
• T+
Total positive counter.
• P+
Partial positive counter, resettable to zero.
• T-
Total negative counter.
• P-
Partial negative counter, resettable to zero.
• NET
Net counter: difference between T+ and T-.
7.3.2 MENU
COUNTERS
Various counters with the
MAG
converter are available: two positive (total and partial) and two negative
(total and partial).
Only partial counters can be set to zero.
To set the counters to zero from the
MAG
software, access the screen “READ” and press the button “ZERO”
(see 1, Fig. 90) that refers to the partial counter that you want to reset.
The same screen also displays a linear bar (see 2, Fig. 90) which shows the percentage of the instantaneous
flow rate in relation to the full scale capacity set in the converter.
The same screen also displays in a special box (see 3, Fig. 90) of the converter: the current “Date” and
“Time.” In an additional module, external temperature or system operating pressure data are available
3
1
2
Fig. 90