![B meters MAG Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/b-meters/mag/mag_instruction-manual_3072628053.webp)
53
MAG
User Manual v2.0
7. MAG CONVERTER PROGRAMMING
7.1 CONVERTER CONFIGURATION METHODS
The configuration of the
MAG
converter can be performed in three different ways:
The MCE-link software can be downloaded from our website www.bmeters.com accessing the
download area. It is also available on the USB flash drive included in the packaging.
7.1.1 Software MCE configuration
Install the
MCE-link
software on the PC, run the program, connect the unit to the PC. In case you are using
a battery-powered sensor, the activation should be completed as follows:
Activation of the MAG in battery mode:
Normal operation of the
MAG
is “sleep mode.” This mode preserves battery life while ensuring the
functional operation of the product. To recall the unit for the next programming, use the magnetic key
(see 1, Fig. 83) provided only with the
MAG
and slide it vertically on the front panel of the converter, to
the point marked with: “ACTIVATE” (Fig. 84). As an alternative, push one of the 4 configuration buttons.
Configuration Buttons
USB-MODBUS Cable
IrCOM Cable
1) Using 4 buttons positioned on
the front display of the converter
(see 1, Fig. 80), unscrewing the
front panel glass (see 2, Fig. 80).
2) From the PC using the RS485
MODBUS output (see electrical
connections) and the
configuration software
MCE-link
.
The USB cable adapter (Fig. 81) is
available from the manufacturer
on request.
3) From the PC via the infra-red
port located on the front of the
converter, above the display and
the
MCE-link
software
configuration program using the
IrCOM cable (Fig. 82) available
from the manufacturer on
request.
Fig. 80
Fig. 81
Fig. 82
1
2
1
Activation Magnet MAG
Activation of the MAG
1
Fig. 83
Fig. 84