Active transistor output powered version
The connected instrument (PLC/external pulse counter) is a passive
digital input which accepts a 24V voltage. It DOES NOT have to
provide ANY additional voltage. The internal 24Vdc power supply is
used to supply necessary voltage.
[Voltage 24Vdc; maximum electrical current 30mA - See “NOTE 1”].
1
12
34
56
78
+
+
-
-
+ -
24V
5.3.2 Programmable output
Passive transistor output
Instrument connected to active input
The connected instrument (PLC/external pulse counter) is an active
digital input which supplies the voltage required to detect the pulses.
The
MAG
acts as a digital switch (logical level).
[Maximum voltage: 5-30Vdc; maximum electrical current 50mA].
1
12
34
56
78
-
+
Instrument connected to passive input
The connected instrument is a passive digital input which accepts
the voltage provided by external power supply in use.
[Voltage5-30Vdc; maximum electrical current 50mA].
1
12
34
56
78
+
-
5-30Vdc
+
-
Active transistor output
Active transistor output (MAG starting from version 6-3)
The connected instrument is a passive digital input which accepts
a 24V voltage. It DOES NOT have to provide ANY additional voltage.
The internal 24Vdc power supply is used to provide necessary
voltage.
[Voltage 24Vdc maximum electrical current 30mA – See “NOTE 1”].
1
12
34
56
78
+ -
+
-
+ -
24V