Unit Elements
4
4 - 6
Perfusor® compact, 2.1 gb
4.3
Syringe Holder
Designation
Ord. No.
Syringe holder, complete . . . . . . . . . . . . . . . . . . . . . . . 3450 6608
with screw and cap
Exchange
1.
Pierce through the cap and remove.
2.
Fasten syringe holder with pin punch.
3.
Remove screw.
4.
Pull off holder.
5.
Push spacer washer 0.8x4x3.2 on the shaft if the syringe is
not recognized.
6.
Insert new syringe holder.
7.
Fit new screw (not the old one) and safety lock with Loctite
274.
8.
Replace new cap.
4.4
Unit Feet
Designation
Ord. No.
Unit feet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 6640
Note
The feet can be turned and used once again. Pull feet out and turn
around or exchange.
Fig.: 4 - 5
Содержание Perfusor compact
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...