Procedural Instructions on the TSC
8
8 - 2
Perfusor® compact, 2.1 gb
Functional Inspection
Switch on unit
1.
Switch on Perfusor and keep ON-button pressed for max. 20
sec. Check the screen display during this time. A device alarm
is triggered if the button is actuated for more than 20 sec.
2.
The following information appears on-screen when the but-
ton is released:
111.
1
222.
2
555.
5
AA
Reference to the instructions
for use (hard- und soft-
ware group)
Last syringe type
3.
An audible alarm sounds three times.
4.
Open lock (drive head).
Check push-button sensor alarm. The piston rod symbol must
flash on the LC-display if a syringe was not inserted.
5.
Insert current step gauge in the OPS 50 ml slot and close sy-
ringe holder.
The lock must engage automatically and the syringe symbol
in the LC display must not flash.
6.
Confirm current step gauge as OPS with F-button. Program as
OPS syringe type beforehand if necessary.
7.
Start pump with a delivery rate of 12.3 ml/h (key sequence 1
2 3).
The pump delivers. The delivery rate set must be displayed.
8.
Open lock.
Alarm by buzzer and positive locking sensor alarm. Drive
stops. Lock must not engage in upper position.
9.
Plug in MFC service connector and actuate the START button.
Drive delivers at 12.3 ml/h.
10. Change delivery rate to 96 ml/h (key sequence C 9 6 F) during
infusion.
The pump delivers. The delivery rate set must be displayed.
Fig.: 8 - 1 Current step gauge
Содержание Perfusor compact
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...