Perfusor® compact, 2.1 gb
B - 1
B
B -
Modification Instructions for Syringe
Holder
Description
The syringe holder must be modified as recognition of syringes
with a large outer diameter by the light barrier system was diffi-
cult. This modification is limited to installation of a washer.
Modification
1.
Pierce through the cap and remove.
2.
Fasten syringe holder with pin punch.
3.
Remove screw.
4.
Pull off holder.
5.
Push washer 0.8x4x3.2 on the shaft.
6.
Insert syringe holder.
7.
Fit new screw (not the old one) and safety lock with Loctite
274.
8.
Place on cap.
9.
Check syringe recognition for proper function
Fig.: B - 1
B - 2
Содержание Perfusor compact
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...