Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 3
Open unit
1.
Loosen 5 screws from the bottom.
2.
Open housing carefully, then
3.
Pull off the ribbon cable from the E-Module and the connec-
tion cable from the motor. Hold the white board holder on the
E-Module when disconnecting!
4.
Dismount both housing halves.
Always check A-Module before replacing the board.
Other modules can only be exchanged without danger of conse-
quential damage if there is no overvoltage.
Connect mains cable when the housing is open. Measure voltage
parallel with capacitor C3. The set value is 6.2 to 6.8 volt.
Fig.: 4 - 1
Ribbon cable to drive
board
Motor cable
Fig.: 4 - 2 A-Module board (up to serial number 38100)
Содержание Perfusor compact
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...