Service Program
3
Perfusor® compact, 2.1 gb
3 - 7
3.
Write
(F7)
The modified values must be loaded in the Perfusor® compact
after calibration, modification of data or the serial number
was input. The status displays „SNr“, „DAbg“ and „LAbg“ must
be ticked. Writing of data is acknowledged by „Writing com-
pleted successfully“. Save the modified data with Menu / File
and store on a storage medium if necessary.
Calibration Menu
1.
Serial Number
(F4)
Enter the serial number when the E-Module is exchanged as
otherwise the EEPROM cannot be written (Dianet type =
1200).
2.
Length Calibration
(F5)
The position of the prealarm light barriers and the drive end
is determined by length calibration. The motor steps deter-
mined are displayed after calibration is terminated.
3.
Pressure Calibration
(F6)
The motor parameters for setting the 3 pressure stages and
the correct switch-off in bolus mode is determined by pres-
sure calibration.
4.
Overload Check
The dynamic pressure test is used to determine whether the
unit was damaged after having been dropped, due to a shock
or impact or when the drive head was dismounted. The drive
must build-up a pressure of
≥
1.6 bar, and the positive locking
sensor must not open.
5.
Parameters
Displays the calibration parameters.
Operation Menu
1.
Service Values
The service values are displayed. These values cannot be
changed. When the default data was specified the service
values are set to zero.
Fig.: 3 - 6
Fig.: 3 - 7
Содержание Perfusor compact
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...