Perfusor® compact, 2.1 gb
8 - 1
8
8 -
Procedural Instructions on the TSC
Visual Inspection
Unit, in General
Completeness, external damage, safe fit of the battery compart-
ment cover and syringe table.
Check cleanliness of device. Check labels and readability.
Syringe Fastening
Check function with OPS 50 ml syringe.
(Syringe holder, axial positioner, drive head, clamp, and push-but-
ton sensor)
Membrane Keyboard
Check adhesion, cleanliness and fit.
Battery Compartment and -Contacts
Check state of contacts (tight fit, not bent).
Unit Feet
Check unit feet for completeness and proper fit.
Mains Lead and Connector
Completeness, damage.
MFC Lead and Connector
Completeness, damage.
Holder for Pole Fixation, Side Snap-in Mechanism
Check function.
8 - 6
Содержание Perfusor compact
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes...