123
(2) Select “Go” for the “RJ 0°C” range on the calibration screen, and press [Go] button.
(3) Wait about 30 seconds, then press the [Go] button.
(4) After finishing the adjustment, the calibration screen for all ranges will be displayed.
(5) Press [ESC] key, and the MENU screen will be displayed.
Handling Precautions
・
If a mistake is made in the course of adjustment, or if some other problem occurs, try to adjust again.
・
To restore the calibration setting for a particular item to the factory default, select “CLR” and press
[ENTER] key.