![Avery Weigh-Tronix FLI 225 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/avery-weigh-tronix/fli-225/fli-225_service-manual_3030602046.webp)
46
FLSC 05 w/FLI 225 Instrument Service Manual
9.1.4 Zero
1.
Place all deadload (pallets, chains, etc.) on the forks.
2.
Select
Zero
from the
Calibrate
screen.
3.
Remove all weight from the scale (except deadload) and press the
Enter
softkey.
The screen will change and may briefly show a weight value other than
zero while processing.
Figure 9.14 Zero in Process
The zero value is stored after processing.
4.
Press the
Enter
or
Esc
softkey to return to the
Calibrate
screen when
prompted.
9.1.5 Span
1.
Select
Span
from the
Calibrate
menu.
2.
Using the data entry procedure, scroll in the calibration weight if other than the
default.
Figure 9.15 Key in Calibration Weight
The Cal Weight value (span weight value) is the known weight you will use when
prompted to “apply load” during calibration.