User manual
Version: v4.5.2
Figure 10.1: Example of an atnet session to upgrade the application firmware.
10.3
Updating the firmware of the fusionTrack shock detection unit
When performing step 7, the fusionTrack
must
be at room
temperature, and switched off for at least 4 h.
The update of the fusionTrack shock detection unit firmware is performed through the command line
software
atnet
. The connection between the fusionTrack and
atnet
is assumed to be established as
presented in Section 10.1. The update procedure is the following:
1. Use the command ‘
update_shockSensor
’ to upload the shock detection unit firmware in the fusionTrack
(see Figure 10.2). For example:
1
update_shockSensor * "C:\ Users \john.doe\ Download \ fTk_accMega_0 .0.1.09. bin"
Listing 10.2: Uploading the new shock sensor detection unit firmware to the fusionTrack.
2. The device will beep twice;
3. Wait for
--> OK
to be displayed (See Figure 10.2).
4. Boot the device in application by running the ‘
reset * 1
’ command;
5. Exit
atnet
with the ‘
exit
’ command;
6. Reboot the fusionTrack and connect with
atnet
;
7. Run the ‘
init_uC *
’ command;
8. Run the ‘
info_uC *
’ command, which should output
--> Calibration=94 Error=0 Status=3
uC_Firmware=...
, followed by
--> OK
.
Atracsys / 2020-06-17 / 16:48:00