In order to clear this warning, an accuracy verification must be performed (See Chapter 13). It can be
done either on site using the Atracsys accuracy verification tool or the device can be sent back to Atracsys
(See Section 1.6). If done on site, the AVT user manual describes the procedure.
18.1.4
FTK_WAR_SHOCK_SENSOR_OFFLINE warning
If the
ftkGetLastFrame
function returns the
ftkError::FTK_WAR_SHOCK_SENSOR_OFFLINE
warning, it means
that the fusionTrack shock sensor was offline. A shock might have happened during the downtime of the
shock sensor.
Depending on the fusionTrack device version, the shock sensor might be not available. Please contact
Atracsys for more information.
In order to determine whether a harming shock occurred during the shock sensor downtime, an accu-
racy verification must be performed (See Chapter 13).
18.1.5
FTK_WAR_SHOCK_SENSOR_AUTO_ENABLE warning
If the
ftkGetLastFrame
function returns the
ftkError::FTK_WAR_SHOCK_AUTO_ENABLE
warning, it means
that the fusionTrack shock sensor was offline for a moment and automatically enabled back at boot. A
shock might have happened during the downtime of the shock sensor.
Depending on the fusionTrack device version, the shock sensor might be not available. Please contact
Atracsys for more information.
In order to clear this warning, the
enable_uC
command of the
atnet64
software is used. In order to
determine whether a harming shock occurred during the shock sensor downtime, an accuracy verification
must be performed (See Chapter 13).
Atracsys / 2020-06-17 / 16:48:00
94 / 113