User manual
Version: v4.5.2
The fusionTrack device is composed of metallic parts and
electronic components and, even if complying with medical
norms, it may interfere or be affected when used in special
environments like Magnetic Resonance Imaging or devices
generating X-Rays. If you intend to operate the fusionTrack
in such environments, please contact Atracsys.
The fusionTrack device is a precision instrument and has to
be handled with greatest care. If it receives a shock, falls on
any surface or is transported without adequate packaging,
it can easily be damaged or decalibrated.
Since the fusionTrack will be integrated in a medical system,
integrator has to conduct the tests related to the electrical
security and electromagnetic compatibility for the whole sys-
tem.
The extension port (10) is not intended for user connections.
Do not connect anything on this plug.
The used power supply must be a limited power source. Any
used power supply must not be able to deliver more than
100 W.
The integrator has to implement a mean to isolate the fusion-
Track electrically from the supply mains.
The fusionTrack device should be recharged for 24 h, if it is
stored for 6 months.
The fusionTrack must
not
be used covered: for instance it
should not be operated under a sterile cover. This would
change the operating temperature of the device and may
change the optical path, both resulting in a change in the
calibration parameters, causing a degradation of the device
accuracy.
Since the fusionTrack can neither be covered with a sterile
drape nor sterilised, the fusionTrack should not be located
in the sterile area.