An active fiducial
marker
is able to transmit its individual
parameters and data to the system. It can also notify the
fusionTrack with alert messages. An error message alert-
ing that battery power is low is sent. That message must
be handled by the integrator to alert the operator that the
marker
is not considered as operational due to a low battery
and the
fiducials
must be turned off.
A weak optical signal from the IR-LEDs can lead to a wrong
estimation of the position. The integrator should ensure the
fiducials are turned off if the battery voltage goes under a
given threshold.
Markers
have to meet the IEC60601-1 requirement regard-
ing ‘Protection against excessive temperatures and other
hazards’.
The integrator should use redundant
fiducials
(i.e. four or
more
fiducials
instead of only three) and require the recon-
structed
marker
to use all of them.
The default parameters have been chosen to ensure optimal
tracking performances. Changing them might impact the
performances of the fusionTrack system.
The fusionTrack device must be cleaned as described in
Chapter 17.
The certification of the fusionTrack has been conducted ex-
cluding the usage of an
aCn
.
The GUI demo software is not part of the fusionTrack soft-
ware system. It is only meant to provide a showcase of what
can be achieved using the SDK. The GUI demo software
does not belong to the fusionTrack product.
The reference implementation is the distributed C / C++ API.
The other language bindings are not subject to torough test-
ing, as the C / C++ do.
When performing step 7, the fusionTrack
must
be at room
temperature, and switched off for at least 4 h.
Atracsys / 2020-06-17 / 16:48:00