Maintenance
In this chapter are found the instructions for maintaining basic safety and performances for the expected
service life, in particular it is explained how to clean the fusionTrack device and its accessories.
17.1
Cleaning the fusionTrack device
The end user manual instruction shall provide a validated
cleaning and disinfection method to the end-user.
The fusionTrack can be disinfected with regular hospital disinfectant detergent applied using a clean
cloth. The optical parts must be cleaned only using camera lenses cleaning solutions. Eyeglass lens tissues
must not be used as it may scratch the lens. Other surfaces must be dried using a clean and dry cloth.
The device must be turned off and disconnected prior to
cleaning. The disinfecting detergent must not be spilled di-
rectly onto the device or any of its components.
It must be ensured that no fluids enter the fusionTrack. This
may damage the fusionTrack by causing short-circuits and
corrode the material.
Only non-corrosive and non-acidic detergents, whose inter-
actions with material are known, may be used.
17.2
Sterilisation of the markers
The integrator must ensure markers are not deformed after
sterilisation.
The battery of an active marker must be properly removed
and disposed of before sterilisation.