This method must be implemented by the integrator and be
automatically run at the application startup and inform the
user on how to proceed.
Any liquid on lenses such as body fluids ejection or conden-
sation could lead to wrong measurements. Lenses have to
be cleaned as soon as they get in contact with liquids. The
zeroing method has to be re-performed after each cleaning.
Some disposable passive
fiducials
cannot be cleaned, as the
reflective coating would be damaged, and may require to be
changed during the operation. The zeroing method has to
be re-performed after changing any
fiducials
.
Accuracy verification must be implemented by the integra-
tor. They can optionnaly purchase the required software and
hardware at Atracsys.
The operator performing accuracy verification must be cor-
rectly trained.
Use of the fusionTrack adjacent to, or stacked with other
equipment should be avoided because it could result in im-
proper operation and can lead to wrong measurements to
serious injury including death. If such use is necessary, the
fusionTrack and the other equipment should be observed to
verify that they are operating normally.
Use of accessories, transducers and cables other than those
specified or provided by Atracsys could result in increased
electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation
and/or loss of performance. Such failure can lead to wrong
measurements, serious injury including death.
Portable radio frequency communications equipment (in-
cluding peripherals such as antenna cables and external an-
tennas) should be used no closer than 30 cm (12 inches)
to any part of the fusionTrack, including cables specified by
Atracsys. Otherwise, degradation of the performance of this
equipment could result, which can lead to wrong measure-
ments, serious injury including death.