User manual
Version: v4.5.2
Markers
or
fiducials
geometrically deformed due to shocks
or manufacturing defects can create wrong measurements
leading to patient injury. Instructions about not using ge-
ometrically deformed
markers
or
fiducials
must be clearly
specified in the end user manual.
Markers
or
fiducials
too close from lenses might lead to
wrong measurement (see Table 2.1). The integrator has
to mention in the end user manual that
markers
or
fiducials
should not be positioned too close from the lenses.
The integrator has to ensure generating an error message
when the
markers
are placed on vibrating instruments and
the frequency of vibrations could affect the measurements.
At all time, the integrator software should monitor the regis-
tration error of the reconstructed
markers
, to prevent badly
reconstructed data to be used.
Instructions about using a calibrated
marker
and not geo-
metrically deformed have to be clearly specified by the inte-
grator to the user.
For single-use
markers
, the end user manual must clearly
state that the
markers
should not be reprocessed. The
markers
packaging must be marked with a label indicating
they cannot be reprocessed.
For single-use
fiducials
, instructions about replacing new
fiducials
after every (single) use must be clearly specified by
the integrator in the end user manual. Packages of passive
fiducials
have to be marked with a label indicating that they
cannot be reprocessed.
A passive fiducial could be damaged due to a shock /
scratch of the reflecting material. Instructions about not us-
ing deteriorated
fiducials
and how to replace a deteriorated
passive
fiducial
must be clearly specified by the integrator in
the end user manual.
Integrator has to specify in the end user manual instructions
how to screw / plug
fiducials
on
markers
.
Atracsys / 2020-06-17 / 16:48:00
105 / 113