Operating manual SUNTEST CPS / CPS+
- 12 -
•
The specimens
&
are placed on the specimen table
%
. Radiation
$
is generated with a Xenon lamp
#
and filtered through an optical filter "coated quartz fil-
ter segment"
"
.
NOTE:
There are optional optical filters (see Chapter 5 "Acces-
sories") available for various spectral distributions (see
chapter. 4.3, "Spectral distribution with additional filters").
•
The UV component of the radiation that is directed
upwards is reflected on the specimens
&
by the mir-
rors
!
placed above the Xenon lamp
#
.
•
The components of the non-dangerous IR radiation (heat
radiation) might penetrate through the mirrors. The
remaining heat is removed out of the unit via the air
circuit of the Xenon lamp.
4
Description of functions
Figure 5
!
Coolant air supply Xenon lamp and
regulation of
test chamber temperature
"
Coolant air outlet Xenon lamp and
regulation of test chamber temperature
4.1
Radiation and Filtering
4.2
Ventilation circuits
•
The coolant air of the Xenon lamp and the test cham-
ber
!
is aspirated on the rear side of the unit (see
chapter 3.3, "Rear side of unit") through an air filter,
flows through the test chamber and the assembly com-
partment
of the irradiance unit and exits through the
exhaust openings
"(
at the left and the rear side of
the unit (see chapter 3.3, "Rear side of unit").
NOTE:
During continuous operation, the hot air emitted by the air
cooling system results in a continuous change in the cli-
mate inside the chamber.
Figure 6
!
Mirrors (UV and light mirrors)
"
Optical filter "coated quartz filter segment"
#
Xenon lamp
$
Radiation
%
Specimen table
&
Specimens
Содержание SUNTEST CPS
Страница 1: ...SUNTEST CPS CPS Operating Manual...
Страница 17: ...17 Operating manual SUNTEST CPS CPS 5 Accessories...