- 23 -
Operating manual SUNTEST CPS / CPS+
Figure19
7
Operation and Shutdown
!
Switch for igniting the Xenon lamp
"
Xenon lamp operating hour meter
#
Rotary switch for adjusting the irradiance
7.4
Operation of SUNTEST CPS
The SUNTEST CPS has a sensor for measuring irra-
diance in the range of 300 - 800 nm. Irradiance in the
unit is regulated to a constant value.
It is possible to measure the actual value by using
the optional sensors for measuring irradiance (see
chapter 5, "Accessories").
1. Use the rotary switch
#
to adjust the irradiance.
Irradiance can be infinitely adjusted in the range of
250 - 765 W/m².
NOTE!
The optional sensors „Radialux“ and „XenoCal” can be used
to measure and display irradiance and radiation.
2. Set the ON(I) / OFF(0) switch
!
to the "ON (I)" posi-
tion. The Xenon lamp is ignited. The duration for which
the Xenon lamp has been on is displayed in the Xenon
lamp operating hour meter
"
.
NOTE!
Replace the Xenon lamp every 1500 operating hours.
!
Two-line display for operator guidance and data
"
Cursor keys for parameter input
#
Keypad for program input
$
Start key, stop key
Figure 20
7.5
Operation of SUNTEST CPS+
program controller
The program controller controls and monitors the-
process of the test programs. The settings are called
via the menu guidance on the control panel. The
menu guidance is described in the accompany-ing
software documentation.
The program controller provides:
•
six test programs,
•
with six test cycles each.
The end of the program can be selected optionally,
according to the criteria:
•
Testing time (time),
•
Irradiance (dose).
Additional program functions are:
•
Programming the light and dark phases,
•
Inputting the irradiance,
•
Inputting the black standard temperature
The display functions are:
•
Displaying the age of lamp and unit,
•
Displaying the current testing parameters in a target/
actual comparison,
•
Displaying the actual value of the test chamber tempe-
rature.