- 9 -
Operating manual SUNTEST CPS / CPS+
2.3
Location requirements
Climatic conditions required for the place of instal-
lation:
•
During continuous operation, the hot air emitted by the
air-cooling system results in a constantly changing cli-
mate in the room. Therefore, the SUNTEST unit should
only be installed in air-conditioned, sufficiently ventila-
ted places of installation.
•
The places of installation must be kept free of dust.
•
The room temperature must be maintained between
18°C and 25°C.
•
The max. relative humidity should be 70%.
The requirement for coolant air is as follows:
•
120 m³/h for lamp cooling and cooling of the test cham-
ber.
2
Installation of unit
2.4
Installation
Installation SUNTEST CPS / CPS+:
• Dimensions of unit (width x depth x height):
780 mm x 350 mm x 350 mm,
•
Weight of unit approx. 30 kg,
•
The unit must be set up on a sufficiently stable labora-
tory bench with a non-flammable support underneath;
the laboratory bench should be aligned so that it is
horizontal.
Caution - Overheating of the unit:
Blocking of the air inlet and air outlets can result in
overheating and damage to the unit!
When installing the unit, the following minimum di-
stances to the neighboring walls or objects must be
maintained:
to the front:
700 mm
to the back:
500 mm
to the top:
350 mm
to both sides:
300 mm