6
Commissioning
!
Xenon lamp
"
Lamp socket 2x
#
Flat connector 2x
$
Plug-in contact
2x
%
Mounting 2x
Figure 14
!
Retaining clip
"
Light mirror (mirror coating)
#
UV mirror (transparent)
$
Holes for retaining clips
%
Mirror support
Figure 15
6.2.5 Installation of the mirrors
WARNING - UV radiation being emitted:
If the mirrors (figure 15) are not used, UV radiation is
emitted on the upper side of the unit and can result
in damage to skin and retina. Always mount the mir-
rors.
NOTE!
The mirrors are pre-mounted. When cleaning or replacing,
install the mirrors in the sequence described.
Clean the mirrors with a soft leather cloth or a brush before
installation.
1. First, place the transparent UV mirrors
#
in the sup-
port
%
.
2. Then position the light mirror
"
.
3. Fix the mirrors in the holes
$
by inserting the two
retaining clips
!
.
4. Insert the mirror support
%
with components fitted into
the radiation compartment.
6.2.4 Installation of the Xenon lamp
Before installation, the Xenon lamp should be cleaned with
a starch-free cloth that is moistened with a small amount
of spirit.
NOTE!
The lamp holder is located in the assembly compartment.
1. Carefully press the Xenon lamp
!
at the lamp sok-
kets
"
into the mounting
%
.
2. Insert the flat connector
#
into the plug-in contact
$
.