37
10-2-4
Performing User Calibration
The variation (deviation)
will be displayed and a
"beep" will be emitted.
Check the value.
Recommended variation:
within 10.0
If the variation exceeds
10.0, refer to p.39.
Spindle Preparation (
C
p.14)
Temperature Sensor
Preparation (
C
p.15)
Sample Preparation (
C
p.15)
Beaker Setup (
C
p.16)
20.0
°
C
--User Calibr.--
Spindle: A1L
Viscosity:
1767.00mPas
Start
20.0
°
C
--User Calibr.--
Attach the
spindle.
OK
20.0
°
C
--User Calibr.--
Calibrating…
Torque: 64.0%
20.0
°
C
--User Calibr.--
Deviation:6.85
Settings saved.
Enter Quit
20.0
°
C
--User Calibr.--
Settings were
not saved.
Exit setup?
Enter Quit
20.0
°
C
--Calibration---
User Calibr.
Reset
[Calibration Menu Screen]
20.0
°
C
--User Calibr.--
Spindle: A1L
Viscosity:
1767.00mPas
Start
Save
Exit without
saving
[Calibration Screen]
[Calibration Screen]
Prepare the spindle,
temperature sensor (only
when using the L Beaker)
and standard liquid.
Beep
N
When an error occurs,
"Calibration error." will
be displayed and
"beeping" tones will be
emitted.
Cannot be selected