User Calibration
The instrument can be calibrated with standard liquid (1 point calibration). Select one
standard liquid from among the following: JS200, JS500, JS1000, or JS2000 (refer to
"Optional Accessories and Replacement Parts,"
C
p.41).
If abnormal measurement values occur, perform the following checks as illustrated below.
10-1
Troubleshooting Tips
See "Abnormal Measurement Values" on p.39.
Measure standard liquid (
C
p.13)
N
Check the viscosity of the standard liquid by using the temperature conversion
chart included with the standard liquid as reference.
Viscosity will differ, depending on the temperature.
Within accuracy range→Normal
Outside accuracy range
↓
Perform user calibration (
C
p.35)
↓
Measure standard liquid (
C
p.13)
Within accuracy range→Normal
Outside accuracy range
↓
Restore default calibration settings (
C
p.38)
↓
Measure standard liquid (
C
p.13)
Within accuracy range→Normal
Outside accuracy range
↓
Contact ATAGO for service and repair (
C
p.44)