![ATAGO VISCO Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/atago/visco/visco_quick-manual_2995845021.webp)
16
3-2-4
Beaker Setup
1.
Attach the beaker to the stand.
m
Firmly attach the beaker, ensuring
there are no spaces or gaps between
the beaker and the stand.
If using the S Beaker:
Attach the small volume beaker adapter to
the stand. Then, attach the beaker.
m
Firmly attach the small volume
beaker adapter and beaker, ensuring
there are no spaces or gaps
between them and the stand.
2.
Place the main unit on the stand.
Check that the surface of the sample is
level with the spindle's reference line.
m
Be sure that the spindle and
temperature sensor do not come in
contact with the beaker.
m
Make sure that no air bubbles are
trapped in the sample.
If air bubbles are present, let the
sample sit and settle in order to
remove the air bubbles.
Quick Measurements (accuracy not guaranteed)
Quick and easy measurements can be taken by placing
the main unit directly on the L Beaker.
m
Quick measurements can only be taken with the
L Beaker.
m
Pour some sample in the beaker up to the quick
measurement reference line.
Spindle name
Reference line (dotted line)
For quick measurement use.
Reference line
Align the beaker
with the groove.
Содержание VISCO