36
10-2-2
Spindle/Beaker Settings
Refer to the "Maximum Measurement Values Guideline Chart" and select the most suitable
spindle and beaker for your application (
C
from p.42).
10-2-3
Standard Liquid Viscosity Settings
Check the viscosity of the standard liquid by using the temperature conversion chart
included with the standard liquid as reference.
Viscosity will differ, depending on the temperature.
20.0
°
C
1767.0
1 2 3 4 5 6 7 8
9 0 . ← →
Back
Enter Quit
20.0
°
C
--User Calibr.--
Spindle: A1L
Viscosity:
1767.00mPas
Start
[Calibration Screen]
←→: use the arrows to
move the cursor.
Back: select "Back" to
delete the value to the
left of the cursor.
Input field
Slide
to select a
value, then push
to input the value.
Confirm
N
Select "Quit" to cancel settings and
return to the calibration screen.
N
Values up to 2 decimal places can
be input.
20.0
°
C
--User Calibr.--
Spindle: A1L
Viscosity:
1767.00mPas
Start
[Calibration Screen]
Blinking
Confirm
Select one of the items listed below:
Item Spindle
Beaker
A1S
A1
S:15mL
A1L
A1
L:100mL
A2S
A2
S:15mL
A2L
A2
L:100mL
A3S
A3
S:15mL
A3L
A3
L:100mL
Содержание VISCO