35
10-2
User Calibration
m
•
Ensure that the main unit, spindle, temperature sensor, beaker and standard
liquid are fully acclimated to the ambient temperature before performing
calibration.
10-2-1
Display
[Setup Screen (setup 2/2)]
20.0
°
C
---Setup 2/2----
Moving Average
Calibration
20.0
°
C
--Calibration---
User Calibr.
Reset
20.0
°
C
--User Calibr.--
Spindle: A1L
Viscosity:
1767.00mPas
Start
[Calibration Menu Screen]
[Calibration Screen]
Spindle/beaker setting
Standard liquid viscosity
setting
Slide
to select an
item, then push
to confirm the selection.
User calibration
Restore Default Calibration
Settings (
C
p.38)
(example display)