OPTIONS
36
MODEL 4500D OPTIONS
Introduction
This section contains a list and descriptions of available factory installed options at the
time of this printing.
Option Label
If your instrument has been modified with options, there will be an option label on the rear
panel of the unit. The option label contains an option(s) code that may be cross-referenced
to the Options List.
For example your options code would appear as follows:
fitted with option 05
OPT: 05
fitted with option 04 and 05
OPT: 0405
4500D Options
Option List
Code Description
04
Arc Detection
05
Floating Return
06
Remote Memory Select
Description
04 Arc Detection
This option provides the user with the capability of detecting low level arcing. The
arc detection circuit has nine preset sensitivity levels.
Arc Detection Specifications
FAILURE SETTINGS
Arc for AC mode (Option 04 only)
Sensitivity range : 1-9
Arc detection : minimum duration < 10
µ
s
SETUP PROCEDURE
To activate Arc Detection and set the sensitivity level.
Please press the SET key until the display shows: