OPERATION
26
OPERATION INSTRUCTIONS FOR Hypot II
Model 4500D
1. POWER UP:
Check that the correct input line voltage has been
selected on the rear panel, either 115 volts AC or 230
volts AC. Plug the line cord into its socket on the rear
panel of the instrument. Connect the male end of the plug
to the outlet receptacle.
Please be sure that the safety ground on the power line cord is not defeated and
that you are connecting to a grounded power source. Also, connect the rear panel
chassis ground for additional safety.
Turn on the POWER switch located on the lower left-hand side of the front panel.
Upon powering the instrument up a POWER ON SELF TEST (POST) will be
automatically performed. This test will check for the condition of the ram chips,
PCB’
s and other critical components. In addition, the display will show the
following message with the model and version number
.
The instrument will recall the last setup that was active and the display will show
the parameters that were programmed into the unit. The instrument is now ready
for operation.
2. SETUP PROCEDURE:
2.1.
To Select the Memory location
Please press the SET key until the display shows
Memory = X
Use the Up/Down arrow keys to select memory locations 1 through 5. After
selecting the memory, press the SET key to view the settings that have been
recalled from memory or to make any changes to these settings.
To store any changes to the settings press the Exit key. The parameters are
automatically updated into the memory location that was selected, when exiting to
the test mode.
CAUTION
ARI HYPOT II
4500D VER: X.X
Current software version
displayed here