![Associated Research Hypot II 4500D Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/associated-research/hypot-ii-4500d/hypot-ii-4500d_operation-and-service-manual_2989500032.webp)
OPERATION
28
2.6.
To set the Dwell Time
Please press the SET key until the display shows:
Timer = X.X s
Use the Up/Down Arrow keys to enter the dwell time setting, then press the
EXIT key to exit to the test mode or toggle to another setting using the SET key.
If the dwell time is set to “0”, the instrument will operate in a continuous ON
mode when the test button is depressed and released. It will stop when the DUT
(Device Under Test) goes into failure or the manual reset button is pressed. The
minimum dwell time setting is 0.2 seconds with 0.1 second resolution.
2.7.
To setup the Ground Continuity Check
Please press the SET key until the display shows:
Continuity=ON
or
Continuity=OFF
Press the Up/Down Arrow key to select the “ON” or “OFF” function of the
Ground Continuity Check mode, the display will show the set status immediately.
Press the EXIT key to exit to the test mode or toggle to another setting using the
SET key.
If the Continuity Function is set to “ON,” the Ground Continuity cable has to be
connected between the “Cont. Check” connector on the instrument panel and the
chassis of the DUT. The instrument will test the ground connection of the DUT
automatically. If continuity is present a green indicator will be illuminated on the
TEST switch and the operator may proceed with the hipot test. The Hypot II will
continue to test the ground connection of the DUT during other tests. If no
continuity is present while the operator tries to initiate a test, the instrument will
signal a continuity failure.
2.8.
To set the Frequency 50/60 Hz
Please press the SET key until the display shows:
Freq =60 HZ
or
Freq =50 HZ
Press the Up or Down Arrow keys to toggle the frequency selection; 50 or 60
hertz, then press the EXIT key to exit to the test mode or toggle to another setting
using the SET key.