OPERATION
33
4. CONNECTION OF REMOTE I/O
Two 9 pin “D” type connectors are mounted on the rear panel which provide
REMOTE-INPUT-OUTPUT control and information. These connectors mate with
standard 9 pin D-subminiature connector provided by the user. The output mates to
a male (plug) connector while the input mates to a female (receptacle) connector.
For best performance, a shielded cable should be used. To avoid ground loops the
shield should not be grounded at both ends of the cable. Suggested AMP part
numbers for interconnecting to the Remote I/O are shown below.
205204-4 PLUG SHELL WITH GROUND INDENTS
205203-3 RECEPTACLE
SHELL
745254-7
CRIMP SNAP-IN PIN CONTACT (for plug)
745253-7
CRIMP SNAP-IN SOCKET CONTACT (for receptacle)
745171-1
SHIELDED CABLE CLAMP (for either plug or receptacle)
747784-3 JACKSCREW SET (2)
Remote Interface Rear Panel
:
4.1.
Signal Outputs on Remote I/O
The rear panel connector provides three output signals to remotely monitor PASS,
FAIL, and PROCESSING conditions. The monitoring signals are provided by
three normally open internal relays, that switch on to indicate the current condition
of the tester. These are normally open free contacts and will not provide any
voltage or current. The ratings of the contacts are 1A / 250 VAC ( 0.5 ADC ). The
signal outputs are provided on the 9 pin female type D connector. Below is a
listing that indicates what conditions activate each pin. When a terminal becomes
active the relay closes thereby allowing the external voltage to operate an external
device.