INSTALLATION
19
REAR PANEL CONTROLS
1.
REMOTE INTERFACE SIGNAL OUTPUT:
9 pin D subminiature female
connector which provides; Pass, Fail and Test in process signals
2.
REMOTE INTERFACE SIGNAL INPUT:
9 pin D subminiature male connector
which provides remote control access over Test, Reset and Interlock.
3.
THERMAL FAN:
To cool the instrument.
4.
HIGH VOLTAGE OUTPUT JACK:
For the connection of the detachable 5 foot
(1.52 m) red high voltage test lead. The silicone rubber insulation is flexible for easy
handling and is rated at 30KVDC. The jack is recessed for safety when this lead is
not being used.
5.
CONTINUITY CHECK OUTPUT JACK:
For the connection of the detachable 5
foot (1.52 m) black return test lead or three prong receptacle adapter box. This jack
is always used when performing a continuity test. Please refer to page 35 for details
on connecting the adapter box between Hypot II and the device under test.
6.
RETURN OUTPUT JACK:
For the connection of the detachable 5 foot (1.52 m)
black return test lead. This lead is always used when performing a test.
7.
INPUT POWER SWITCH:
Line voltage selection is set by the position of the
switch. In the down position it is set for 115 volt operation, in the up position it is set
for 230 volt operation.
8.
CHASSIS GROUND (EARTH) TERMINAL:
This safety terminal should be
connected to a good earth ground before operation
9.
INPUT POWER RECEPTACLE:
Standard IEC 320 connector for connection to
a standard NEMA style line power (mains) cord.
Содержание Hypot II 4500D
Страница 5: ...SECTION 1 OPERATORS MANUAL ...
Страница 43: ...SECTION 2 SERVICE MANUAL ...