8.2 Fault diagnostics
The EFT / BURST generator PEFT Junior was designed and produced in accordance with the
quality standard ISO 9001.
If however a functional failure does occur, it is important to complete the following table which will
assist in the search for the problem and help reduce system down time.
Symptom Questions
The PEFT Junior does not power
on and no lights are illuminated !
Is the PEFT Junior plugged in ?
Is the supply within the limits specified in
1.4 ?
The generator fuses are intact ?
Is the generator mains switch selected to
ON ?
Nothing is displayed on the P90
front panel and no lamps are
illuminated !
Is the supply within the limits specified in
1.4 ?
Has the PEFT Junior been subjected to
mechanical shocks during transport (see
point 8.2) ?
The message:
generator fatal error
is displayed !
Switch off the PEFT Junior and wait 5
minutes, when powered on again does the
same error message occur ?
The message:
generator fatal error
is displayed again !
Note the full error message and refer to
section 8.3
Tab.
8.2