![artisan PEFT Junior Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/artisan/peft-junior/peft-junior_user-manual_2978433013.webp)
4
EFT/ Burst Generator PEFT Junior
1.2.4 Coupling modes
The EFT / burst can be delivered as follows.
Coupling into control lines using
the capacitive coupling clamp
To HV-OUT
Direct output to SHV
connector
Coupling into single phase mains
To-LINE 1PHASE
Output coupled to
integrated single
phase filter
Coupling into three phase mains
To-LINE 3PHASE
Direct output to SHV
connector,
with control of the
external three phase
filter
Tab.1.2.4
1.2.5 Single phase integrated filter
Maximum operating voltage
Ueff
250 V
Maximum operational current
DC
max.10 A
50 Hz max. 16 A
400 Hz max. 9 A
Bandwidth
1 to 100 MHz
Coupling attenuation
< 2 dB
Asymetric filter attenuation
> 20 dB
Tab.
1.2.5
1.2.6 Connections
High voltage output
to coaxial connector type SHV
Mains input to integrated filter
connector type IEC 320 250 V / 16 A
Mains output from integrated filter
Euro (Schuko) 250 V / 16 A