EFT/ Burst Generator PEFT Junior
51
The following error messages, can be accessed from the P90 using the option "Remote Control
Debug":
Error Message
Meaning
input buffer ovfl
The input buffer memory has exceeded 100
characters
timeout occured
timeout exceeded in listen mode
header >4 characters
Header exceeds 4 characters
option required
the option part of a command is missing
unknown header
command not recognised
invalid argument
parameter incorrect
timeout while talk
timeout exceeded in talk mode
no query here
this is not an interrogation command
query expected
waiting for an interrogation
not valid in local
command not allowed in local mode
not valid while run
command only valid in standby mode
Tab.
6.3.5.3c
6.3.5.4 Error control
A list of options specific to the PEFT Junior is given in 6.3.5.5.
The options available for systems equiped with the P90 are given below.
ERR
(ERRor query)
The response to a command ERR?, is
a number
Code
Description
0
no error
1
command only valid in remote mode
2
unknown command
3
argument not valid
4
no query available
5
command only valid in standby mode
8
timeout during transmit
16
parity error during RS-232C transfer
32
overflow in receive buffer
64
other failure
The codes 1 to 5 are applicable to the
immediately preceeding command. The
error signal is reset after each
interrogation.
Tab.
6.3.5.4a
examples Interrogation
command
response
CHV 1.25;ERR?<EOS>
0
CHV &%;ERR?<EOS>
3