46
EFT/ Burst Generator PEFT Junior
HAEFELY
P90.1
CONTROL UNIT
F1
F2
F3
F4
remote control
setup menu
INTERFACE IEEE-488
LOAD SETUP -->
The parameter menu for IEEE-488 is active.
Press to define the parameters.
Interface menu
HAEFELY
P90.1
CONTROL UNIT
F1
F2
F3
F4
BUS ADDRESS 12
→
EOS CR+LF
After entry of all the parameters, press to
return to the interface menu. Press F4 to
save this configuration and continue
Configuration
menu IEEE-488
Configuration parameters for the RS-232C and IEEE-488 are pre defined. It is not possible to
select other values to those contained in the communication file.
6.3.3 PC Configuration
The control device or PC must be configured to be compatible with the sytem communication
requirements.
Example 1: Using the HP48SX controller, and serial communication.
HP48SX
P90
From the "I/O setup" menu
IR/WIRE WIRE
BAUDRATE
9600
ASCII/BINARY ASCII
DATABIT 8
BAUD 9600
STOPBIT
1
PARITY
NONE 0
PARITY
NONE
CHEKSUM TYPE 3
PROTOCOL
XON/XOFF
TRANSLATE CODE 1
EOS
CR+LF
Tab.
6.3.3a
Example 2: Using the HP300 controller communicating by IEEE.
HP9000
series300
P90
HP-IB configuration
BUS ADRESS
12
EOS
CR+LF
Tab.
6.3.3b
6.3.4 Using the RS232/IEEE type 730 interface
The Remote-Control type 730 interface contains two communication standards, IEEE-488 and
RS-232C.